Toluene micro flow measurement
email: porter@microflowmeter.com Shanghai JSN Micro Flow MeterCO.,LTD is a manufactory integrating scientific research, production, sales andservice. It is located in the Shanghai Industrial Zone. The company is based onthe people-oriented principle and takes the program as the degree. Competitive,product covers a variety of materials and super micro liquid flow meter,ovalgear flow meter ,mass flow meter.spur gear flow meter,gear pump,gear meteringpump etc instrumentation instrument. Widely applicable to trade settlement andmeasurement in various industries such as petroleum, chemical, environmentalprotection, metallurgy, electric power, transportation, food, medicine,military industry, scientific research, etc.

Toluene micro flow measurement
At presentdomestic Oval gear flow meter structure is complex, bulky, bulky, noisy, smallscale, not suitable for low viscosity liquid, the market is more common withDTO series and OGM series. For these problems the country has been unable toovercome, while the Shanghai JiShen instrument Instrument company oval gear Ovalgear flow meter does not have the above problems, such as the same DN25caliber, our model CX-M25-AL, flow range: 0.5 ~ 100L / M, range of 1: 200 orhigher, and weighs only 1.5KG , While the DTO-25, the flow of 600-6000L / H,range ratio of only 1:10, the weight of up to 23KG. All this due totechnological breakthroughs, our chief engineer Tu workers, he is a formerSiemens senior engineer, the use of equipment It is also Siemens CNC MachiningCenter with cutting accuracy up to 0.001mm, which ensures that we produce highquality flow meters.
Shanghai JSN Micro Flow Meter Co., Ltd. is the second manufacturer ofmicro Oval gear flow meter in China. It is developed by Tu engineer (formerSiemens senior engineer) with independent intellectual property rights. Theproduction and processing equipment fully adopts Siemens CNC machining centerto ensure the minimum flow 0.1ml / min, which is rare in the world, theaccuracy of over 90% of peers. We can do more temperature 200C, the pressurecan be customized to 1000bar, many customers use it, but also unbelievable.Also domestic Most of the flowmeter use a year after the new calibration, butthe base depth meter after hundreds of conditions of medium test, the servicelife of more than 8 years, and without secondary calibration, since theestablishment of the base depth instrument Instrument determined to beworld-class flow meter, breaking the monopoly of international brands in themirco flow meter industry, so that customers can buy in the country realGerman-made quality flow meter.

Toluene micro flow measurement
The Ovalgear flow meter is a volumetric flow meter that consists of a flow chamber in achamber and one or several elements that are rotated by fluid flow to form aflow meter. Leakage between the rotating element and the inner wall and theselected working range The internal flow rate is negligible. The rotation ofthe element is transmitted mechanically or by other means to the indicatingdevice to indicate the volume flow of the fluid flowing through.

Toluene micro flow measurement
Technicaldata of Gear type flow meter
Model | K-factor | Minimum flow | Flow range | Accuracy |
G0.025 | 0.025ml | 0.001L/Min | 0.008~2L/Min | 0.5% |
G0.04 | 0.04ml | 0.004L/Min | 0.2~4 L/min | 0.5% |
G0.1 | 0.4ml | 0.008L/Min | 0.01~8L/min | 0.5% |
G0.2 | 0.245ml | 0.01L/Min | 0.16~16 L/min | 0.5% |
G0.4 | 0.4ml | 0.01L/Min | 0.16~16 L/min | 0.5% |
G1 | 1.036ml | 0.02L/Min | 0.4~80 L/min | 0.5% |
What is the reason why the tiny gauge flow meter is larger than the actual value? the reason:1. The flow has a large pulse.2. The medium is mixed with gas.3. Medium viscosity is too large.Treatment measures:1. Reduce the pulsation of the flow in the pipeline.2. Install degasifier.3. Recalibrate, replace the adjustment gear pair and correct it.